The FDA has designated a 3 year transition period, after which the previous version will not be accepted- REC#19-8 (IEC 60601-1-2:ed4.0 + ANSI AAMI IEC 60601-1-2:2014).
New testing requirements include
- Testing at minimum and maximum input voltage levels for conducted emissions and voltage dips and short interruptions
- Power frequency magnetic field at either 50 or 60 Hz so long as the frequency is the same as that used to power the equipment
- A new technical reference standard has been added IEC 61000-4-39:2017
- Magnetic field immunity testing requirements from 9 kHz to 13.56 MHz